FASTASeq 300 Desktop High-Throughput Sequencer GeneMind

GeneMind
#fastaseq-300

FASTASeq 300 is a desktop sequencing platform designed for targeted sequencing and low-pass whole-genome sequencing. Its chemistry, flow cell design, fluidics and base identification algorithms are optimized to deliver consistent data quality with practical run-time flexibility.

Key advantages

  • Supports several flow cell types and multiple read lengths
  • Pre-configured plug-and-play cartridges
  • RFID reader module support
  • Libraries can be used directly after preparation
  • Listed rapid workflows include SE50 mNGS in 4 hours and PE150 targeted sequencing in 13 hours

For Research Use Only. Not intended for diagnostic or clinical procedures. Availability may vary by country.

Type: Sequencing system

Section: Sequencers

lwh: 684x644x615 mm

Weight: 145000 g

Product Advantages
Specification
Run Modes
Application Capacity

Product Advantages

AdvantageDetails
FlexibleSupports two flow cell types and multiple read lengths; supports Multi-Time Output; libraries can be loaded automatically or manually by flow cell lane.
Easy-to-usePre-configured plug-and-play cartridges; RFID reader module support; prepared libraries can be used directly for sequencing.
RapidListed workflows include mNGS SE50 in 4 hours, targeted sequencing PE75 in 9.5 hours and targeted sequencing PE150 in 13 hours.

Specification

ParameterValue
Flow cell optionsFCL, FCM, FCH, FCP, FCX
Maximum listed output150 Gb with FCP PE150; 60 Gb with FCX PE300
Q30>= 90% for listed short-read modes; >= 85% for PE300
Dimensions684 mm x 644 mm x 615 mm
Weight145 kg
Operating power2000 VA
Power requirements100-240 VAC, 50/60 Hz
Operating environment19-25 deg C; 20-80% RH, non-condensing; altitude 0-3000 m
Control computerIntel Xeon Silver 4126; 32 GB DDR4 x 6; 480 GB SSD; 2 TB HDD; Windows 10 x64

Run Modes

Flow cellLaneReads/FCRead lengthOutputQ30TAT
FCL250MSE1005 Gb>=90%5.5 hr
FCM4125MSE759.4 Gb>=90%4.5 hr
FCM4125MPE7518.8 Gb>=90%9.5 hr
FCM4125MPE15037.5 Gb>=90%13 hr
FCH4280MSE7521 Gb>=90%5 hr
FCH4280MPE7542 Gb>=90%9.5 hr
FCH4280MPE15084 Gb>=90%16 hr
FCP4500MSE7537.5 Gb>=90%7.5 hr
FCP4500MPE7575 Gb>=90%14 hr
FCP4500MPE150150 Gb>=90%24 hr
FCX2100MSE40040 Gb>=80%31.5 hr
FCX2100MPE30060 Gb>=85%46.5 hr

Application Capacity

ApplicationData requirementFCMFCHFCP
NIPT Plus>7M unique reads/sample122448
PGT-A>3.5M unique reads/sample123264
TB-tNGS>2M raw reads/sample3696192
Small tumor panel, tissue0.2 Gb/sample>48>96>192
Small tumor panel, ctDNA5 Gb/sample61632
mNGS20M reads/sample41224